Suggestion for improvement:
The DAQ Hardware wiki page says "Voltage supplied over the DC jack should be between 5V and 12V".
This might be misleading for users that are unfamiliar with the system, since we regularly come across situations in which users plug in 9V or 12V with a standard setup and fry their miniscopes.
For the miniscope v4 perhaps the range should not be as extense, or maybe a more conservative procedure can be suggested:
Voltage supplied over the DC jack depends on setup configuration such as cable length, type of cable, commutation, etc. A good starting point would be a 6V voltage, and likely not more than 9V is needed since the required voltage at the miniscope is 5V. We suggest you use a variable power supply and closely monitor how much power is actually reaching the scope to avoid damage.
Suggestion for improvement:
The DAQ Hardware wiki page says "Voltage supplied over the DC jack should be between 5V and 12V".
This might be misleading for users that are unfamiliar with the system, since we regularly come across situations in which users plug in 9V or 12V with a standard setup and fry their miniscopes.
For the miniscope v4 perhaps the range should not be as extense, or maybe a more conservative procedure can be suggested:
Voltage supplied over the DC jack depends on setup configuration such as cable length, type of cable, commutation, etc. A good starting point would be a 6V voltage, and likely not more than 9V is needed since the required voltage at the miniscope is 5V. We suggest you use a variable power supply and closely monitor how much power is actually reaching the scope to avoid damage.