Expand Up @@ -141,13 +141,24 @@ print(metadata)
```
```output
OME(
experimenters=[{'id': 'Experimenter:0', 'user_name': 'Zeiss'}],
instruments=[<1 field_type>],
images=[<1 field_type>],
plates=[<1 field_type>],
experimenters=[{'id': 'Experimenter:Zeiss', 'user_name': 'Zeiss'}],
instruments=[<1 field_type>],
images=[<2 field_type>],
structured_annotations={
'xml_annotations': [
{
'description': 'ZEN 2012 (blue edition)1.1.1.0',
'id': 'urn:lsid:allencell.org:Annotation:AcquisitionSoftware',
'value': {},
'kind': 'xmlannotation'
}
]
}
)
```

In this case, we can see it is split into three categories:
In this case, we can see it is split into various categories including
`experimenters`, `images` and `instruments`. You can look inside each of these
using `.category` e.g.:
```python
Expand All @@ -174,7 +185,7 @@ processing steps don't overwrite this original image.
Explore the metadata in the console to answer the following
questions:

- What type of microscope was used to take this image?
- Which manufacturer made the microscope used to take this image?
- What detector was used to take this image?
- What does each channel show? For example, which fluorophore is used? What is
its excitation and emission wavelength? (hint: look inside the image's
Expand All @@ -184,10 +195,10 @@ pixel metadata: `metadata.images[0].pixels`)

## Solution

### Microscope model
### Microscope manufacturer

Under `metadata.instruments[0].microscope`, we can see that an 'Axio Imager Z2'
microscope was used.
Under `metadata.instruments[0].microscope`, we can see that the microscope
manufacturer was Zeiss.

### Detector
Under `metadata.instruments[0].detectors`, we can see
Expand All @@ -196,18 +207,18 @@ that this used an HDCamC10600-30B (ORCA-R2) detector.
### Channels

Under `metadata.images[0].pixels.channels`, we can see one entry per channel
- `Channel:0:0`, `Channel:0:1` and `Channel:0:2`.
- `Channel:1:0`, `Channel:2:0` and `Channel:3:0`.

In the first (`metadata.images[0].pixels.channels[0]`), we can see that its
`name` is TagYFP, a fluorophore with `emission_wavelength` of 524nm and
`fluor` is TagYFP, a fluorophore with `emission_wavelength` of 524nm and
`excitation_wavelength` of 508nm.

In the first (`metadata.images[0].pixels.channels[1]`), we can see that its
`name` is mRFP1.2, a fluorophore with `emission_wavelength` of 612nm and
In the second (`metadata.images[0].pixels.channels[1]`), we can see that its
`fluor` is mRFP1.2, a fluorophore with `emission_wavelength` of 612nm and
`excitation_wavelength` of 590nm.

In the first (`metadata.images[0].pixels.channels[2]`), we see that its name
is `TL DIC` and no `emission_wavelength` or `excitation_wavelength` is listed.
In the last (`metadata.images[0].pixels.channels[2]`), we see that no `fluor`,
`emission_wavelength` or `excitation_wavelength` is listed.
Its `illumination_type` is listed as 'transmitted', so this is simply an image
of the yeast cells with no fluorophores used.

Expand Down