* Add incident review for the July 2026 live clinic gateway failure First document in docs/incidents/. Covers the first production use of the gateway imaging flow (6-7 July): three stacked faults with one identical symptom - a private DNS blackhole via a broken-but-Approved workers, and a zombie relay listener - private endpoint, the asyncio websockets client failing under gevent plus what was done, the improvements that followed (phase-tagged error reporting, runbooks, echo probe, end-to-end test participant, healthcheck and per-site toggle work), and the learnings. * Couple of extra points in the incident review
1 parent 217f9bd commit fa5a29c

1 file changed

Lines changed: 38 additions & 0 deletions

File tree

Lines changed: 38 additions & 0 deletions
Original file line numberDiff line numberDiff line change
@@ -0,0 +1,38 @@
1+
# Incident review: Hull live clinic gateway failure, 6–7 July 2026
2+
3+
**Severity**: No patient impact. The clinic ran to completion on the manual-images fallback. Gateway (automatic) imaging was unavailable for the first live clinic; full end-to-end success was achieved the same afternoon.
4+
5+
## Summary
6+
The first production use of the gateway imaging flow failed: worklist items sent from Rubie never reached the hospital gateway. Diagnosis was unusually hard because three independent faults were stacked, all producing the same symptom (a 30-second timeout on every send), and the system’s error reporting could not distinguish them. Each fault was peeled off in turn over ~36 hours; the full pipeline – check-in → worklist → modality → images → upload → Rubie – worked in production for the first time on the afternoon of 7 July.
7+
8+
## Symptoms
9+
- Every worklist send from prod Rubie failed with a generic “Timeout waiting for response from gateway”; nothing appeared on the modality worklist.
10+
- All infrastructure looked healthy: relay listener connected (count 1), managed identity and role assignments correct, gateway services running, config verified repeatedly on both sides.
11+
- Independent probes succeeded where the app failed (a laptop-based relay echo round-tripped in 0.4s), deepening the confusion.
12+
13+
## Root causes
14+
- **Private DNS blackhole**: the relay hostname resolved, inside the prod VNet, to the relay namespace’s private endpoint, which was silently broken despite showing *Approved* (later attributed to an Azure-side fault or provisioning race at PE creation; resolved weeks later by deleting and recreating the PE, which has worked since).
15+
- **Runtime incompatibility**: the asyncio websockets client cannot reliably open connections inside gevent-patched gunicorn workers (DNS resolution via asyncio's thread executor degrades to 10s+, exhausting the connection budget). Every environment where sends had ever succeeded bypassed gevent (local dev, shell probes) – the production execution model had never actually been tested.
16+
- **Silent listener death**: overnight, the gateway's relay connection died without the process noticing or the relay deregistering it – a “zombie” listener that looked connected while receiving nothing.
17+
18+
**Contributing factors**: relay send failures were recorded only in a database field, not logged; a single exception handler conflated “connection never opened” with “gateway never replied”; VM telemetry export was broken throughout, forcing diagnosis via remote commands; there was no admin visibility of gateway actions at the start of the incident.
19+
20+
## What we did
21+
- Shipped diagnostic capability during the incident: a GatewayActions admin page.
22+
- Built a standalone relay echo probe to test the relay → gateway path with Rubie removed; its container-side variant exposed the DNS/PE fault.
23+
- Fixed DNS (interim public CNAME; later superseded by PE recreation), replaced the asyncio client with the synchronous websockets client ([Rubie ADR-009](https://github.com/NHSDigital/dtos-manage-breast-screening/blob/main/docs/adr/ADR-009-Synchronous_websockets_client_for_relay_sends.md)), and adopted a pre-clinic listener restart to clear zombie connections.
24+
- Invoked the designed fallback: the clinic ran entirely on manual image recording, invisibly to participants.
25+
- Relay token refresh implemented
26+
27+
## Improvements made since
28+
- **Error reporting**: every relay failure now names its failing phase in both the admin and logs; unknown actions get structured error replies instead of silent connection closes.
29+
- **Runbooks**: [clinic preparation](../deployment/runbooks/clinic-preparation.md) checklist; [live-clinic debugging](../deployment/runbooks/live-clinic-debugging.md); onboarding/promotion updates; diagnostic toolkit ships with the gateway release.
30+
- **Automated end-to-end test participant**: a synthetic test can now be sent through the real production pipeline on demand – Rubie dispatches a test worklist item over the relay; the gateway stores it and invokes its modality emulator in place of the physical machine, which queries the worklist, generates DICOM images, and submits them through the standard PACS-and-upload path back to Rubie, where the round trip is confirmed against a dedicated test-action record (kept separate from clinical data).
31+
- **Rehearsal practice**: successful remote dummy run completed ahead of the next live clinic.
32+
- **In flight**: automated per-gateway healthcheck (health proven by traffic, surfaced in admin, wired into deploy pipelines); per-site imaging toggle so switching a site to manual is an audited admin action rather than an emergency deploy or record deletion; gateway listener liveness/token-renewal fixes.
33+
34+
## Learnings
35+
- Health must be proven by traffic, not state. Listener counts, PE connection status, and service states each individually misled us during this incident. Everything we’ve built since sends a real message and judges the reply.
36+
- The fallback earned its keep. Manual-images mode meant zero patient impact and removed all time pressure from diagnosis.
37+
- Institutional knowledge belongs in the repo. Every finding was folded into version-controlled runbooks the same week, several of which were corrected again as evidence improved.
38+
- We had no production gateway we could actually touch. Every diagnostic step on the gateway side ran through Arc run-commands against a hospital-owned machine – serially queued, minutes per round trip, dependent on hospital IT for RDP and power, and twice found switched off entirely. A team-operated gateway VM in the production environment (Azure-hosted, Arc-onboarded, ring0, connected to the prod relay and API but to no modality) would have let us reproduce and inspect the failures directly under real prod identity and networking.