Requirements

Those requirements are based on the analysis of different commercially available sensors as well as medical literature around ventilation. We made an effort to distinguish between what would be an ideal and what would be still acceptable for the use case.

Key Requirements

  • Proximal Flow measurement
  • Proximal Pressure measurement
  • Compatibility with standard connectors used in BVMs and ventilators (see ISO 5356-1:2015)
  • Affordability (manufacturing cost <10$)
  • Ease of manufacturing (only 3D printing and common parts)

Functional Parameters

Parameter Minimum Viable Value Ideal Value
Pressure Range 0 - 2 kPa (0 - 20 cmH2O) -4 - +4 kPa (-42 - +41 cmH2O)
Pressure Resolution 0.2 kPa (2 cmH2O) 0.1 kPa (1 cmH2O)
Pressure Accuracy <10% m.v <5% m.v
Pressure Noise Level <5% m.v <2.5% m.v
Pressure Update Time <100 ms <10 ms
Flow Range 0 - 100 slm -250 - +250 slm
Flow Resolution 1 slm 0.2 slm
Flow Offset 2 slm 0.2 slm
Flow Accuracy <20% m.v <5% m.v
Flow Noise Level <5% m.v <2.5% m.v
Flow Update Time <100 ms <10 ms
Pressure drop <1 kPa (10 cmH2O) <0.5 kPa (5 cmH2O)
Dead Space <25 mL <10 mL

m.v: measured value
slm: standard liter per minute

Operating Parameters

Parameter Minimum Viable Value Ideal Value
Operating Temperature +5 - +35 °C 0 - +50 °C
Operating Humidity (relative) 10 - 95 % (non-condensing) 10 - 95 % (non-condensing)
Media Compatibility Air, N2, O2 Air, N2, O2, other non- aggressive gases