File tree Expand file tree Collapse file tree 1 file changed +49
-0
lines changed
Expand file tree Collapse file tree 1 file changed +49
-0
lines changed Original file line number Diff line number Diff line change @@ -151,6 +151,55 @@ \section{Demo}
151151\thickhline
152152\end {tabularx }
153153
154+ \section {Batronix }
155+
156+ A driver for Batronix Magnova oscilloscope is available in the codebase.
157+ It has been developped, with the help of Batronix team, according to official SCPI documentation for Magnova oscilloscope
158+ (Magnova® SCPI Manual - Version 1.5). This driver has been developed and tested on a Magnova BMO350 running firmware version 1.7.1.
159+
160+ Here are the known limitations of this driver (that should be fixed with further Magnova firmware updates):
161+ \begin {itemize }
162+
163+ \item For now, Magnova's SCPI implementation does not support digital signal acquisition.
164+
165+ \item Selecting specific Arbitrary waveform from function generator is not possible with current Magnova SCPI implementation.
166+
167+ \item Time Scale / Memory Depth / Sample Rate synchronization only work if the scope is in "Extended Capture" mode (Acquire menu).
168+
169+ \end {itemize }
170+
171+ \begin {tabularx }{16cm}{lllX}
172+ \thickhline
173+ \textbf {Device Family } & \textbf {Driver } & \textbf {Transport } & \textbf {Notes } \\
174+ \thickhline
175+ Magnova BMO & Batronix & lan & Connection path: <IP>:5025 \\
176+ \thickhline
177+ \end {tabularx }
178+
179+ \subsubsection {Typical Performance (Magnova BMO350, LAN) }
180+
181+ \begin {tabularx }{16cm}{llX}
182+ \thickhline
183+ \textbf {Channels } & \textbf {Memory depth } & \textbf {WFM/s }\\
184+ \thickhline
185+ 1 & 20K & 9.9 \\
186+ \thinhline
187+ 2 & 10K & 6.6 \\
188+ \thinhline
189+ 4 & 12.5K & 3.99 \\
190+ \thinhline
191+ 1 & 100K & 9.7 \\
192+ \thinhline
193+ 4 & 125K & 3.97 \\
194+ \thinhline
195+ 1 & 5M & 4.5 \\
196+ \thinhline
197+ 4 & 5M & 1.26 \\
198+ \thickhline
199+ \end {tabularx }
200+
201+ These figures were obtained from a Magnova BMO350 running firmware version 1.7.1.
202+
154203\section {Digilent }
155204
156205Digilent oscilloscopes using the WaveForms SDK are all supported using the `` digilent" driver in libscopehal. This
You can’t perform that action at this time.
0 commit comments