# Key AAPS features
2
2
3
-
###Loop mode
3
+
## Loop mode
4
4
5
5
The loop status is shown on the main screen with one of the icons below.
6
6
@@ -12,7 +12,7 @@ See [AAPS Screens > The Homescreen > Loop status](#AapsScreens-loop-status) for
12
12
13
13
(KeyAapsFeatures-OpenLoop)=
14
14
15
-
####Open Loop
15
+
### Open Loop
16
16
17
17
**AAPS** continuously evaluates all available data (IOB, COB, BG...) and makes treatment suggestions (temporary basal rates) on how to adjust your therapy if necessary.
18
18
@@ -22,7 +22,7 @@ This option is for getting to know how **AAPS** works or if you are using an uns
22
22
23
23
(KeyAapsFeatures-LGS)=
24
24
25
-
####Low Glucose Suspend (LGS)
25
+
### Low Glucose Suspend (LGS)
26
26
27
27
In this mode, [maxIOB](#Open-APS-features-maximum-total-iob-openaps-cant-go-over) is set to zero.
28
28
@@ -32,7 +32,7 @@ This mode is available starting at **[Objective 6](#objectives-objective6)**.
32
32
33
33
(KeyAapsFeatures-ClosedLoop)=
34
34
35
-
####Closed Loop
35
+
### Closed Loop
36
36
37
37
**AAPS** continuously evaluates all available data (IOB, COB, BG...) and automatically adjusts the treatment if necessary (*i.e.* without further intervention by you) to reach the set [target range or value](#profile-glucose-targets) (bolus delivery, temporary basal rate, insulin switch-off to avoid hypo etc.).
38
38
@@ -42,7 +42,7 @@ Closed Loop is only possible if you are in **[Objective 7](#objectives-objective
42
42
43
43
(Open-APS-features-autosens)=
44
44
45
-
## Gevoeligheidsdetectie (Autosens)
45
+
## Autosens
46
46
47
47
- Autosens is an algorithm which looks at blood glucose deviations (positive/negative/neutral).
48
48
- Op basis van deze afwijkingen, bepaalt het algoritme hoe insulinegevoelig/resistent je bent.
@@ -170,7 +170,7 @@ If this option is enabled, the sensitivity detection (autosens) can lower the ta
170
170
171
171
This setting is available when one of "Enable dynamic sensitivity" or "Enable Autosens feature" are enabled.
172
172
173
-
### Activeer SMB
173
+
### Enable SMB
174
174
175
175
Enable this to use SMB functionality. If disabled, no **SMBs** will be given.
176
176
@@ -206,7 +206,7 @@ If this setting is enabled, SMB is enabled when there is any temp target set (ea
206
206
207
207
This setting is not visible if "Enable SMB always" is switched on.
208
208
209
-
#### Activeer SMB na koolhydraten
209
+
#### Enable SMB after carbs
210
210
211
211
If enabled, SMB is enabled for 6h after carbohydrates are announced, even if COB has reached 0.
#### Maximum aantal minuten basaal om SMB te limiteren tot
228
+
#### Max minutes of basal to limit SMB to
229
229
230
230
This is an important safety setting. This value determines how much SMB can be given based on the amount of basal insulin in a given time, when it is covered by COBs.
231
231
@@ -243,7 +243,7 @@ Default value : the same as **Max minutes of basal to limit SMB to**.
243
243
244
244
This setting is only visible if "Enable SMB" and "Enable UAM " are switched on.
245
245
246
-
### Activeer UAM
246
+
### Enable UAM
247
247
248
248
With this option enabled, the SMB algorithm can recognize unannounced meals. This is helpful if you forget to tell **AAPS** about your carbs or estimate your carbs wrong and the amount of entered carbs is wrong or if a meal with lots of fat and protein has a longer duration than expected. Without any carb entry, UAM can recognize fast glucose increase caused by carbs, adrenaline, etc., and tries to adjust it with SMBs. This also works the opposite way: if there is a fast glucose decrease, it can stop SMBs earlier.
249
249
@@ -261,7 +261,7 @@ In any case, the required carbs will be displayed in the COB section on your hom
261
261
262
262

263
263
264
-
### Geavanceerde instellingen
264
+
### Advanced Settings
265
265
266
266
You can read more here : [OpenAPS docs](https://openaps.readthedocs.io/en/latest/docs/While%20You%20Wait%20For%20Gear/preferences-and-safety-settings.html).
267
267
@@ -279,13 +279,13 @@ Default value: 4 (shouldn’t be changed unless you really need to and know what
279
279
280
280
(Open-APS-features-advanced-meal-assist-ama)=
281
281
282
-
## Geavanceerde maaltijdhulp (AMA)
282
+
## Advanced Meal Assist (AMA)
283
283
284
284
AMA, the short form of "advanced meal assist" is an OpenAPS feature from 2017 (oref0). OpenAPS Advanced Meal Assist (AMA) allows the system to high-temp more quickly after a meal bolus if you enter carbs reliably.
285
285
286
286
You can find more information in the [OpenAPS documentation](https://newer-docs.readthedocs.io/en/latest/docs/walkthrough/phase-4/advanced-features.html#advanced-meal-assist-or-ama).
287
287
288
-
### Maximale E/uur dat een tijdelijke basaalstand kan toedienen (OpenAPS "max-basal")
288
+
### Max U/hr a Temp Basal can be set to (OpenAPS "max-basal")
289
289
290
290
This safety setting helps **AAPS** from ever being capable of giving a dangerously high basal rate and limits the temp basal rate to x U/h. It is advised to set this to something sensible. A good recommendation is to take the highest basal rate in your profile and multiply it by 4 and at least 3. For example, if the highest basal rate in your profile is 1.0 U/h you could multiply that by 4 to get a value of 4 U/h and set the 4 as your safety parameter.
291
291
@@ -301,7 +301,7 @@ The hardcoded parameters in **AAPS** are:
301
301
302
302
*See also [overview of hard-coded limits](#Open-APS-features-overview-of-hard-coded-limits).*
303
303
304
-
### Max basaal IOB dat OpenAPS kan toedienen\[E\] (OpenAPS "max-iob")
304
+
### Maximum basal IOB OpenAPS can deliver\[U\] (OpenAPS "max-iob")
305
305
306
306
This parameter limits the maximum of basal IOB where **AAPS** still works. If the IOB is higher, it stops giving additional basal insulin until the basal IOB is under the limit.
307
307
@@ -323,7 +323,7 @@ Here, you can choose, if you want to use the [sensitivity detection](../DailyLif
323
323
324
324
If you have this option enabled, autosens can adjust targets (next to basal and ISF), too. This lets **AAPS** work more 'aggressive' or not. The actual target might be reached faster with this.
325
325
326
-
### Geavanceerde instellingen
326
+
### Advanced Settings
327
327
328
328
- Normally you do not have to change the settings in this dialogue!
329
329
- If you want to change them anyway make sure to read about details in [OpenAPS docs](https://openaps.readthedocs.io/en/latest/docs/While%20You%20Wait%20For%20Gear/preferences-and-safety-settings.html#) and to understand what you are doing.