Module 01 simulates a Digital Operating Room.
The applications have been tested to work in Debian-based environments with a GUI, including those in WSL2 with GUI support (Windows 10 and 11), as well as on macOS.
All run commands in this README are launched from the repository root. The project-level launch.py script is the runtime entrypoint; there is no module-local launcher in this folder.
This module implements 5 interactive Connext applications using the MedTech Reference Architecture to simulate a digital operating room. You can find a diagram below:
The Patient Sensor application's primary function is to publish simulated patient vitals data on Topic t/Vitals.
It has no user interface.
The Patient Monitor application's primary function is to subscribe to patient vitals data on Topic t/Vitals.
It displays the received patient vital data in a GUI as a live "Patient Monitor Graph" line plot.
The Arm application's primary function is to simulate a surgical robot with 5 motors: Base, Shoulder, Elbow, Wrist, and Hand. It receives commands on Topic t/MotorControl to control the movement of the individual motors.
It displays the motor angles as an "Arm Motor Angles" line plot. This is a very basic representation of the movement of a robotic arm. We've intentionally made it simple to focus on the data transfer.
The Arm Controller application's primary function is to send motor-movement commands on Topic t/MotorControl.
It presents buttons to administer Arm motor commands, and shows an "Alerts" panel to display observed events.
The Orchestrator application primarily acts as a system application state observer and controller. It detects the presence and current status of other system applications on Topics t/DeviceHeartbeat and t/DeviceStatus, respectively. It can also command device-level actions, such as "Start" or "Shut Down" on Topic t/DeviceCommand.
It displays current device statuses, presents buttons to administer device commands, and shows an "Alerts" panel to display observed events.
Complete the shared setup in the root Quick Start section. That covers prerequisites, environment setup, the project-level build, and security artifact generation.
Module-specific notes:
- If you plan to use secure mode, make sure the security artifacts from the root README have been generated.
- There is no module-local build or launch script; use the project-level
build.pyandlaunch.pyscripts from the repository root.
Important: Run the commands below from the repository root.
launch.pylives at the project root and is the single runtime entrypoint for this project.
Run operating room applications from the repository root:
# From the repository root
python3 launch.py 01-operating-roomTo run with security enabled, use the -s option:
# From the repository root
python3 launch.py 01-operating-room -sThis script does the following:
-
Set
NDDS_QOS_PROFILESenvironment variable to load the appropriate XML configuration files (QoS Profiles, Domains, DomainParticipants). -
Launch the application executables.
Note, applications can be launched individually by name, e.g. python3 launch.py 01-operating-room Arm PatientMonitor. Refer to module.json for the list of available app names and QoS configuration.
Observe and play around with the interactive operating room applications. The following are some ideas to get started:
- From the Orchestrator application, send a "PAUSE" command to the Patient Sensor Medical Device. Observe the effect in the Patient Monitor GUI application. Resume the Patient Sensor by sending a "START" command.
- From the Arm Controller application, send a command to stop all Arm motors. Observe the effect in the Arm GUI application. Resume just the Elbow motor by clicking the respective "PLAY" button in the Arm Controller application. While still stopped, increment or decrement the Wrist motor angle by clicking the respective "+" or "-" buttons in the Arm Controller application. Resume all Arm motors by pressing "PLAY ALL".
- Ungracefully terminate the Arm application by closing the application window. Observe the effect in the Orchestrator application "Alerts" panel.
- Initiate a graceful remote shutdown of the Arm Controller by sending the appropriate command from the Orchestrator. Observe the effect of the Arm Controller application.
Press Ctrl-C in the terminal where launch.py is running to terminate all application processes.
Here are a few hands-on exercises to demonstrate use cases of Connext features in this module.
The module uses a 200 ms Deadline QoS period for DataWriters and DataReaders on the t/DeviceHeartbeat Topic to detect when applications are no longer active. Specifically, the DataWriters and DataReaders use a 200 ms Deadline period. This means if the Orchestrator application does not receive a t/DeviceHeartbeat within 200 ms, it will log an entry, such as the following, to the "Alerts" panel:
DeviceType::PATIENT_MONITOR is no longer sending heartbeats. Updating Status to OFF.
Let's try testing how modifying QoS Profiles can adjust system behavior without modifying any application logic.
-
Kill all active operating room applications.
-
Increase the
<datareader_qos>Deadline period in QoS Profile DataFlow::Heartbeat in Qos.xml to 5 seconds.Note, you should undo this configuration change after completing this demonstration.
<deadline> <period> <sec>5</sec> <nanosec>0</nanosec> </period> </deadline>
-
Relaunch the operating room applications.
-
While monitoring the Orchestrator "Alerts" panel, ungracefully terminate the Patient Monitor application by closing the respective application window.
Observe: You should see the alert mentioned above logged, 5 seconds after the Patient Monitor application has been closed.
-
Undo the QoS change to revert to the recommended configuration.
The module uses Content-Filtered Topics for DataReaders in all applications that subscribe to the t/DeviceCommand Topic. This allows the Orchestrator application to use a single Topic to publish device-specific commands. This allows the subscribing application logic to not have to check if the device command is relevant before processing it.
For example, the Patient Monitor application uses a device = 'PATIENT_MONITOR' filter expression for its DataReader subscribed to the t/DeviceCommand Topic.
Let's try testing how removing a content filter can adjust system behavior without modifying any application logic.
-
Kill all active operating room applications.
-
Comment out the
<content_filter>tag and contained content within it, for dr/DeviceCommand under dp/PatientMonitor in ParticipantLibrary.xml.Note, you should undo this configuration change after completing this demonstration.
-
Relaunch the operating room applications.
-
While monitoring both the Arm Controller and Patient Monitor application windows, send a "Shut Down" command from the Orchestrator application for the Arm Controller application only.
Observe: You should see both the Arm Controller and Patient Monitor applications shut down and exit. This is because the Patient Monitor is no longer configured to receive only device commands for Patient Monitor apps but any app.
-
Undo the Content Filter change to revert to the recommended configuration.